Performance of two different electronic apex locators during the removal of calcium silicate-based sealers

dc.contributor.authorGöçer, Pelin Andaç
dc.contributor.authorDurmazpınar, Parla Meva
dc.contributor.authorBaştürk, Fatima Betül
dc.date.accessioned2024-09-11T20:08:22Z
dc.date.available2024-09-11T20:08:22Z
dc.date.issued2023
dc.departmentİstanbul Gelişim Üniversitesien_US
dc.description.abstractAim: The aim of this in vitro study is to evaluate the accuracy of Root ZX and Propex Pixi during the retreatment of root canals obturated with different calcium silicate-based root canal sealers. Methodology: Root canals of ninety human premolar teeth were shaped with ProTaper universal rotary files up to size F2 under 5.25% NaOCl irrigation for the present study. Root canal filling was carried out by lateral condensation technique using with gutta-percha and sealers; MTA Fillapex, CeraSeal or AH Plus. After seven days, the obturating materials were removed using D-Race files. While actual working length was recorded by direct observation, electronic working lengths were determined by Propex Pixi or Root ZX for tolerance limits of 0.5 mm and 1.0 mm. Data were analyzed using Shapiro Wilk, Mann-Whitney U, and Kruskal-Wallis tests (p < 0.05). Results: The performance of Propex Pixi and Root ZX was found to be statistically similar and did not differ on the root canals filled with different root canal sealers (p < 0.05). Propex Pixi and Root ZX demonstrated more accurate measurements in the range of ±1 mm compared to ±0.5 mm in all groups, regardless of the type of root canal sealer used when filling the root canals before retreatment (p < 0.05). Conclusion: Root ZX and Propex Pixi can be used confidently in the retreatment of the root canals in which CeraSeal, MTA Fillapex, and AH Plus are used as sealersen_US
dc.identifier.doi10.5577/idr.2023.vol13.s1.3
dc.identifier.endpage25en_US
dc.identifier.issn2146-1767
dc.identifier.issueSpecial Issue 1en_US
dc.identifier.startpage20en_US
dc.identifier.trdizinid1238655en_US
dc.identifier.urihttps://doi.org10.5577/idr.2023.vol13.s1.3
dc.identifier.urihttps://search.trdizin.gov.tr/tr/yayin/detay/1238655
dc.identifier.urihttps://hdl.handle.net/11363/8708
dc.identifier.volume13en_US
dc.indekslendigikaynakTR-Dizinen_US
dc.language.isoenen_US
dc.relation.ispartofInternational Dental Researchen_US
dc.relation.publicationcategoryMakale - Ulusal Hakemli Dergi - Kurum Öğretim Elemanıen_US
dc.snmz20240903_Gen_US
dc.subjectApex locatoren_US
dc.subjectapical constrictionen_US
dc.subjectcalcium silicate-based sealeren_US
dc.subjectretreatmenten_US
dc.subjectworking lengthen_US
dc.titlePerformance of two different electronic apex locators during the removal of calcium silicate-based sealersen_US
dc.typeArticleen_US

Dosyalar