dc.description.abstract | Amaç: Bu çalışmanın amacı, Yoğunluk Ayarlı Radyoterapi (YART) planlarının hastaya özel kalite kontrolü
için kullanılan film, in-vivo elektronik portal görüntüleme cihazları (Electronically Portal Imaging DevicesEPID) ve iki boyutlu (2B) detektörlerin hata tespit yeteneklerini karşılaştırmaktır.
Yöntem: Bu çalışmada, Alderson Rando fantom kullanılmıştır. İlk olarak, Alderson Rando fantomun
bilgisayarlı tomografi (BT) görüntüleri üzerine glioblastoma (GBM) ve baş-boyun (HN) kanseri tedavisini
temsil eden iki tedavi planı oluşturulmuştur. Bu planlar, orijinal planlar olarak adlandırılmıştır. Daha sonra,
bilinçli bir şekilde farklı türde hatalar yaratılarak hatalı planlar elde edilmiştir. Film, in-vivo EPID ve 2B
detektörler ile tüm planların hastaya özel kalite kontrolleri yapılmıştır. Elde edilen sonuçlar, gamma analizi
yöntemi kullanılarak karşılaştırılmıştır.
Bulgular: Orijinal planlar tüm yöntemlerde gamma analizi testini geçmiştir. HN radyoterapisi için çok
yapraklı kolimatör (ÇYK) hatası, in-vivo EPID dozimetrisi ile tespit edilmiştir. Cihaz kalibrasyonu ile ilgili
hatalar, tüm yöntemlerde yakalanmıştır. Ancak fraksiyon dozu, gantri ve kolimator açısına bağlı hatalar,
kullanılan yöntemlerle tespit edilememiştir.
Sonuç: Hastaya özel kalite kontrol uygulamalarında kullanılan film, 2B dedektörler ve in-vivo EPID
dozimetresi arasında hata tespit yetenekleri bakımından belirgin bir fark bulunmamaktadır. Tüm yöntemler,
kalibrasyon hatalarını tespit etme konusunda duyarlıdır. ÇYK hatalarının tespiti için in-vivo EPID
dozimetrisi ön plana çıkmaktadır. Ancak özellikle gantri ve kolimatöre ait mekanik hatalar, mevcut
sistemlerle tespit edilememektedir. | en_US |
dc.description.abstract | Aim: The aim of this study is to compare the error detection capabilities of film, in-vivo EPID dosimetry,
and two-dimensional detectors used in patient-specific quality assurance of IMRT plans.
Method: Alderson Rando phantom was used in our study. In the first stage, two treatment plans were
created on CT images of the Alderson Rando phantom. One of the created plans represents GBM treatment,
while the other represents the treatment of the head and neck. These plans are called as the original plan.
Then, the plans with errors were obtained by deliberately creating different types of errors. Quality controls
of all plans were made with film, in-vivo EPID and 2D detectors. The obtained results were compared with
the gamma analysis method.
Results: The original plans passed the gamma analysis for all methods. MLC error was detected by in vivo
EPID dosimetry for head and neck irradiation. Errors related to machine calibration were caught in all
methods. Errors related to fraction dose, gantry and collimator angle could not be determined by the
methods used in the study.
Conclusions: There is no difference in error detection capabilities between film, 2D detectors and in-vivo
EPID dosimeters, which are widely used in patient-specific quality control applications in IMRT treatments.
All methods are sensitive to calibration errors. In-vivo EPID dosimetry comes to the fore in the detection of
MLC errors. However, mechanical errors especially in the gantry and collimator could not be detected with
the existing systems. | en_US |
dc.language.iso | tur | en_US |
dc.publisher | İstanbul Gelişim Üniversitesi Yayınları / Istanbul Gelisim University Press | en_US |
dc.relation.isversionof | https://doi.org/10.38079/igusabder.1199555 | en_US |
dc.rights | info:eu-repo/semantics/openAccess | en_US |
dc.rights | Attribution-NonCommercial-NoDerivs 3.0 United States | * |
dc.rights.uri | http://creativecommons.org/licenses/by-nc-nd/3.0/us/ | * |
dc.subject | Radyoterapi | en_US |
dc.subject | YART | en_US |
dc.subject | hasta QA | en_US |
dc.subject | dozimetri | en_US |
dc.subject | Radiation therapy | en_US |
dc.subject | IMRT | en_US |
dc.subject | patient QA | en_US |
dc.subject | dosimetry | en_US |
dc.title | Yoğunluk Ayarlı Radyoterapide Hastaya Özel Kalite Kontrol Uygulamalarında Kullanılan Gereçlerin Hata Tespit Yeteneklerinin Karşılaştırılması | en_US |
dc.title.alternative | Comparison of Error Detection Capabilities of Equipment Used in Patient-Specific Quality Assurance in Intensity Modulated Radiotherapy | en_US |
dc.type | article | en_US |
dc.relation.ispartof | İstanbul Gelişim Üniversitesi Sağlık Bilimleri Dergisi | en_US |
dc.department | İstanbul Gelişim Üniversitesi | en_US |
dc.authorid | https://orcid.org/0000-0001-9448-8278 | en_US |
dc.identifier.issue | 19 | en_US |
dc.identifier.startpage | 139 | en_US |
dc.identifier.endpage | 147 | en_US |
dc.relation.publicationcategory | Makale - Ulusal Hakemli Dergi - Başka Kurum Yazarı | en_US |